prog saved
This commit is contained in:
@@ -91,9 +91,9 @@ the beam rate was generally less than \SI{8}{\kilo\hertz}.
|
||||
|
||||
To make sure that we will analyse good data, a low level data quality checking
|
||||
was done on the whole data sets. The idea is plotting the variations of basic
|
||||
parameters, such as noise level, length of \tpulseisland{}, \tpulseisland{}
|
||||
rate, time correlation to hits on $\mu$Sc, \ldots on each channel during the
|
||||
data collecting period. Runs with significant difference from the nominal
|
||||
parameters, such as noise level, length of raw waveforms, pulse rate, time
|
||||
correlation to hits on the muon counter on each channel during the data
|
||||
collecting period. Runs with significant difference from the averaging
|
||||
values were further checked for possible causes, and would be discarded if such
|
||||
discrepancy was too large or unaccounted for. Examples of such trend plots are
|
||||
shown in \cref{fig:lldq}.
|
||||
@@ -120,9 +120,10 @@ the data quality checking shown in \cref{fig:lldq}. The data set contains
|
||||
\num{6.43E7} muon events.
|
||||
|
||||
Firstly, the number of charged particles emitted after nuclear muon capture on
|
||||
the active target is calculated. This number then is normalised to the number
|
||||
of nuclear muon capture to obtain an emission rate. Finally, the rate is
|
||||
compared with that from the literature.
|
||||
the active target is calculated. The charged particles yield then is normalised
|
||||
to the number of nuclear muon capture to obtain an emission rate.
|
||||
%Finally, the
|
||||
%rate is compared with that from the literature.
|
||||
|
||||
\subsection{Event selection}
|
||||
\label{sub:event_selection}
|
||||
@@ -196,22 +197,21 @@ starting from at least 1200~ns, therefore another cut is introduced:
|
||||
\end{enumerate}
|
||||
|
||||
The three cuts~\eqref{eqn:sir2_prompt_cut},~\eqref{eqn:sir2_muE_cut} and
|
||||
~\eqref{eqn:sir2_2ndhit_cut} reduce the muon events sample to the size of
|
||||
\num{9.32E+6}.
|
||||
~\eqref{eqn:sir2_2ndhit_cut} reduce the sample to the size of \num{9.82E+5}.
|
||||
|
||||
The number of stopped muons can also be calculated from the number of muonic
|
||||
X-rays recorded by the germanium detector. The X-rays are emitted during the
|
||||
cascading of the muon to the muonic 1S state in the time scale of \SI{E-9}{\s},
|
||||
so the hit caused by the X-rays must be in coincidence with the muon hit on the
|
||||
active target. Therefore an additional timing cut is applied for the germanium
|
||||
hits:
|
||||
detector hits:
|
||||
\begin{equation}
|
||||
\lvert t_{\textrm{Ge}} - t_{\mu\textrm{ counter}} \rvert < \SI{500}{\ns}
|
||||
\label{eqn:sir2_ge_cut}
|
||||
\end{equation}
|
||||
|
||||
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
||||
\subsection{Number of charged particles with energy above \SI{2}{\MeV}}
|
||||
\subsection{Number of charged particles with energy above \SI{3}{\MeV}}
|
||||
\label{sub:number_of_charged_particles_with_energy_from_8_10_mev}
|
||||
As shown in \cref{fig:sir2_1us_slices} and illustrated by MC simulation
|
||||
in \cref{fig:sir2_mc_pdfs}, there are several components in
|
||||
@@ -235,11 +235,11 @@ the energy spectrum recorded by the active target:
|
||||
\centering
|
||||
\includegraphics[width=0.45\textwidth]{figs/sir2_meas_spec}
|
||||
\includegraphics[width=0.45\textwidth]{figs/sir2_mc_pdfs}
|
||||
\caption{The observed spectrum in the timing window 1500 -- 9500~ns (left)
|
||||
\caption{The observed spectrum in the timing window 1300 -- 10000~ns (left)
|
||||
and its components from MC simulation (right). The charged particles
|
||||
spectrum is obtained assuming the spectrum shape and emission rate from
|
||||
Sobottka and Wills~\cite{SobottkaWills.1968}. The relative scale between
|
||||
components is arbitrarily chosen for the purpose of illustration.}
|
||||
Sobottka and Wills~\cite{SobottkaWills.1968}. The relative scales between
|
||||
components are arbitrarily chosen for the purpose of illustration.}
|
||||
\label{fig:sir2_mc_pdfs}
|
||||
\end{figure}
|
||||
|
||||
@@ -270,7 +270,8 @@ in~\ref{fig:sir2_mll_fit}, the yields of backgrounds and signal are:
|
||||
\begin{align}
|
||||
n_{\textrm{beam e}} &= 23756 \pm 581\\
|
||||
n_{\textrm{dio}} &= 111340 \pm 1245\\
|
||||
n_{\textrm{sig}} &= 207201 \pm 856
|
||||
n_{\textrm{sig}} &= 2.57 \pm 856
|
||||
\label{eqn:sir2_n_chargedparticles}
|
||||
\end{align}
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
@@ -282,12 +283,6 @@ in~\ref{fig:sir2_mll_fit}, the yields of backgrounds and signal are:
|
||||
\label{fig:sir2_mll_fit}
|
||||
\end{figure}
|
||||
|
||||
The total number of charged particles from time zero is then calculated to be:
|
||||
\begin{equation}
|
||||
N_{\textrm{charged particles}} =(149.9\pm 0.6)\times 10^4
|
||||
\label{eqn:sir2_Nchargedparticle}
|
||||
\end{equation}
|
||||
|
||||
% subsection number_of_charged_particles_with_energy_from_8_10_mev (end)
|
||||
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
||||
\subsection{Number of nuclear muon captures}
|
||||
@@ -336,92 +331,34 @@ or correction was made.
|
||||
%The area of the $(2p-1s)$ peak is $N_{(2p-1s)} = 2981.5 \pm 65.6$,
|
||||
%obtained by subtracting the background of 101.5 from the spectral integral of
|
||||
%2083 in the region from 396 to 402 keV.
|
||||
The area of the $(2p-1s)$ peak is $2929.7 \pm 56.4$ obtained by fitting
|
||||
a Gaussian peak on top of a first-order polynomial background to the spectrum
|
||||
in \cref{fgi:sir2_xray} in the region from \SIrange{395}{405}{\keV}.
|
||||
%The area of the $(2p-1s)$ peak is $2929.7 \pm 56.4$ obtained by fitting
|
||||
%a Gaussian peak on top of a first-order polynomial background to the spectrum
|
||||
%in \cref{fgi:sir2_xray} in the region from \SIrange{395}{405}{\keV}.
|
||||
Using the same procedure of fitting and correcting described in
|
||||
\cref{sub:germanium_detector}, the number of X-rays is calculated to be 370.
|
||||
Details of the correction factors are given in \cref{tab:sir2_xray_corr}.
|
||||
\begin{table}[htb]
|
||||
\begin{center}
|
||||
\begin{tabular}{l}
|
||||
\toprule
|
||||
\textbf{Col1}\\
|
||||
\midrule
|
||||
item1\\
|
||||
\bottomrule
|
||||
\end{tabular}
|
||||
\end{center}
|
||||
\caption{Corrections for the number of X-rays from the active target.}
|
||||
\label{tab:sir2_xray_corr}
|
||||
\end{table}
|
||||
|
||||
This number of X-rays needs to be corrected for following effects:
|
||||
\begin{itemize}
|
||||
\item Self-absorption effect: the X-rays emitted could be absorbed by the
|
||||
target itself, the probability of self-absorption becomes larger in case of
|
||||
thick sample and low energy photons.
|
||||
For this silicon target of 1500~\si{\um}\ thick and the photon energy of
|
||||
400~keV, and assuming a narrow muon stopping distribution at the centre of
|
||||
the target, the self-absorption correction is estimated to be:
|
||||
\begin{align}
|
||||
k_{\textrm{self absorption}} &= \dfrac{\mu t}{1 - e^{-\mu t}} \nonumber\\
|
||||
&= \dfrac {9.614\times 10^{-2} \times 2.33 \times 0.75 \times 10^{-1}}
|
||||
{1 - e^{-9.614\times 10^{-2} \times 2.33 \times 0.75 \times 10^{-1}}}\nonumber \\
|
||||
%&= \dfrac{1}{0.992} \nonumber\\
|
||||
&= 1.008
|
||||
\end{align}
|
||||
where $t = \SI{0.075}{\cm}$ is the thickness of the target, and $\mu$ is
|
||||
the linear attenuation coefficient of silicon for a photon of 400~keV. The
|
||||
value of $\mu$ is calculated as product of the density of silicon
|
||||
$\rho = \SI{2.33}{\g\per\cm^3}$ and its mass attenuation coefficient
|
||||
$\mu/\rho = \SI{9.614E-2}{\cm^2\per\g}$ taken
|
||||
from the NIST's X-ray Mass Attenuation Coefficients
|
||||
table~\footnote{\url{http://www.nist.gov/pml/data/xraycoef}}.
|
||||
\item Dead time of the germanium detector system: there are two causes of
|
||||
dead time in our germanium detector, (a) the insensitive period due to long
|
||||
pulse time, and (b) the reset pulses of the transistor reset preamplifier.
|
||||
The effects of the two dead time could be calculated by examining the
|
||||
interval between two consecutive pulses on the germanium detector in
|
||||
\cref{fig:sir2_ge_deadtime}.
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
\includegraphics[width=0.85\textwidth]{figs/sir2_ges_self_tdiff}
|
||||
\caption{Interval between to consecutive pulses on the germanium
|
||||
detector. The peak at 57~\si{\us}\ indicates the pulse length, and
|
||||
the bump at about 2000~\si{\us}\ shows the width of the reset
|
||||
pulses. The average count rate of this detector is extracted from the
|
||||
decay constant of the time spectrum to be
|
||||
$5.146 \times 10^{-7}\textrm{ ns}^{-1} = 514.6 \textrm{ s}^{-1}$}
|
||||
\label{fig:sir2_ge_deadtime}
|
||||
\end{figure}
|
||||
|
||||
The correction factor for the pulse length is calculated by the formula:
|
||||
\begin{align}
|
||||
k_{\textrm{pulse length}} &= e^{2\times \textrm{(pulse length)}
|
||||
\times \textrm{(count rate)}} \nonumber\\
|
||||
&= e^{2\times 57\times10^{-6} \times 514.6} \nonumber\\
|
||||
&= 1.06
|
||||
\end{align}
|
||||
The 2-ms-long reset pulses effectively reduce the actual measurement time
|
||||
compared to other channels, so the correction factor for the effect is:
|
||||
\begin{align}
|
||||
k_{\textrm{reset pulse}} &= \frac{\textrm{(measurement time)}}
|
||||
{\textrm{(measurement time)}
|
||||
- \textrm{(number of reset)}\times
|
||||
\textrm{(reset pulse length)}}\nonumber\\
|
||||
&= 1.033
|
||||
\end{align}
|
||||
|
||||
\item The true coincidence summing is negligibly small due to the far
|
||||
geometry as mentioned in the calibration process, so no correction is made.
|
||||
%%TODO
|
||||
\item The geometrical acceptance of the detector: the absolute efficiency
|
||||
calibration was done with a point-like source, but the actual points of
|
||||
origin of the X-rays have a finite spatial distribution. The correction
|
||||
factor is estimated to be \ldots
|
||||
\end{itemize}
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
\includegraphics[width=0.85\textwidth]{figs/ge_eff_mc_finitesize_vs_pointlike}
|
||||
\caption{Ratio between geometrical acceptance of the germanium detector in
|
||||
two cases: point-like source and finite-size source.}
|
||||
\label{fig:ge_eff_mc_finitesize_vs_pointlike}
|
||||
\end{figure}
|
||||
|
||||
The number of X-rays after applying all above corrections is 3293.5. The X-ray
|
||||
intensity in \cref{tab:mucap_pars} was normalised to the number of stopped
|
||||
muons, so the number of stopped muons is:
|
||||
The X-ray intensity in \cref{tab:mucap_pars} was normalised to the number of
|
||||
stopped muons, so the number of stopped muons is:
|
||||
|
||||
\begin{align}
|
||||
N_{\mu\textrm{ stopped}} &=
|
||||
\dfrac{N_{(2p-1s)}}{\epsilon_{2p-1s}\times I_{(2p-1s)}}\nonumber\\
|
||||
&= \dfrac{3293.5}{4.54\times10^{-4} \times 0.803} \\
|
||||
&= 9.03\times10^6 \nonumber
|
||||
&= \dfrac{370}{4.38\times10^{-4} \times 0.803} \\
|
||||
&= 1.05\times10^6 \nonumber
|
||||
\end{align}
|
||||
where $\epsilon_{(2p-1s)}$ is the calibrated absolute efficiency of the
|
||||
detector for the 400.177~keV line in \cref{tab:xray_eff}, and
|
||||
@@ -431,15 +368,15 @@ $I_{(2p-1s)}$ is the probability of emitting this X-ray per stopped muon
|
||||
Taking the statistical uncertainty of the peak area, and systematic
|
||||
uncertainties from parameters of muon capture, the number of stopped muons
|
||||
calculated from the X-ray measurement is
|
||||
$(9.03 \pm 0.31)\times 10^6$. This figure is consistent with the number of
|
||||
stopped muons of $9.32\times10^6$ after the cuts described in the event
|
||||
$(10.50 \pm 0.65)\times 10^5$. This figure is consistent with the number of
|
||||
stopped muons of $9.82\times10^5$ after the cuts described in the event
|
||||
selection process.
|
||||
|
||||
The number of nuclear captured muons is:
|
||||
\begin{equation}
|
||||
N_{\mu\textrm{ nucl.capture}} =
|
||||
N_{\mu\textrm{ stopped}}\times f_{\textrm{cap.Si}}
|
||||
= 9.03 \times 10^6 \times 0.658 = 7.25 \times 10^6
|
||||
= 10.05 \times 10^5 \times 0.658 = 6.91 \times 10^5
|
||||
\label{eqn:sir2_Ncapture}
|
||||
\end{equation}
|
||||
where the $f_{\textrm{cap.Si}}$ is the probability of nuclear capture per
|
||||
@@ -738,7 +675,7 @@ X-rays recorded and the number of captures are shown in
|
||||
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
||||
|
||||
\subsection{Lifetime measurement}
|
||||
\label{sub:lifetime_measurement}
|
||||
\label{sub:lifetime_measurement}w
|
||||
To check the origin of the protons recorded, lifetime measurements were made by
|
||||
cutting on time difference between a hit on one thick silicon and the muon
|
||||
hit. Applying the time cut in 0.5~\si{\us}\ time steps on the proton
|
||||
@@ -762,7 +699,7 @@ before that, the time constants are shorter ($655.9\pm 9.9$ and $731.1\pm8.9$)
|
||||
indicates the contamination from muon captured on material with higher $Z$.
|
||||
Therefore a timing cut from 500~ns is used to select good silicon events, the
|
||||
remaining protons are shown in \cref{fig:si16p_proton_ecut_500nstcut}.
|
||||
The spectra have a low energy cut off at 2.5~MeV because protons with energy
|
||||
The spectra have a low energy cut off at 2.5~MeV because protons with energy:
|
||||
lower than that could not pass through the thin silicon to make the cuts as the
|
||||
range of 2.5~MeV protons in silicon is about 68~\si{\um}.
|
||||
\begin{figure}[htb]
|
||||
|
||||
@@ -58,7 +58,8 @@ bookmarks
|
||||
\RequirePackage[final]{listings}
|
||||
\RequirePackage{xfrac}
|
||||
%% Units
|
||||
\RequirePackage[]{siunitx}
|
||||
%\RequirePackage[]{siunitx}
|
||||
\RequirePackage[detect-weight=true, detect-family=true]{siunitx}
|
||||
\RequirePackage{hepnames}
|
||||
%% Various fonts ...
|
||||
%\RequirePackage[T1]{fontenc}
|
||||
|
||||
Reference in New Issue
Block a user