prog saved
This commit is contained in:
@@ -402,10 +402,11 @@ correlation between detectors would be established in the analysis stage.
|
||||
|
||||
At the beginning of each block, the time counter in each digitiser is reset to
|
||||
ensure time alignment across all modules. The period of 110~ms was chosen to be:
|
||||
{\em i} long enough compares to the time scale of several \si{\micro\second}\ of the
|
||||
physics of interest, {\em ii} short enough so that there is no timer rollover
|
||||
on any digitiser (a FADC runs at its maximum speed of \SI{170}{\mega\hertz} could
|
||||
handle up to about \SI{1.5}{\second} with its 28-bit time counter).
|
||||
{\em i} long enough compared to the time scale of several \si{\micro\second}\
|
||||
of the physics of interest, {\em ii} short enough so that there is no timer
|
||||
rollover on any digitiser (a FADC runs at its maximum speed of
|
||||
\SI{170}{\mega\hertz} could handle up to about \SI{1.5}{\second} with its
|
||||
28-bit time counter).
|
||||
|
||||
To ease the task of handling data, the data collecting period was divided into
|
||||
short runs, each run stopped when the logger had recorded 2 GB of data.
|
||||
@@ -495,8 +496,8 @@ the recorded pulse height spectrum is shown in \cref{fig:ge_eu152_spec}. The
|
||||
source was placed at the target position so that the absolute efficiencies can
|
||||
be calculated. The peak centroids and areas were obtained by fitting a Gaussian
|
||||
peak on top of a first-order polynomial background. The only exception is the
|
||||
\SI{1085.84}{\keV} line because of the interference of \SI{1089.74}{\keV},
|
||||
the two were fitted with two Gaussian peaks on top of a first-order
|
||||
\SI{1085.84}{\keV} line because of the interference of the \SI{1089.74}{\keV}
|
||||
gamma, the two were fitted with two Gaussian peaks on top of a first-order
|
||||
polynomial background.
|
||||
|
||||
The relation between pulse height in ADC value and energy is found to be:
|
||||
@@ -527,30 +528,86 @@ a little worse at 3.1~\si{\keV}~for the annihilation photons at
|
||||
\label{fig:ge_fwhm}
|
||||
\end{figure}
|
||||
|
||||
The absolute efficiencies of the referenced points, and calculated
|
||||
efficiencies at the X-ray of interest are presented in
|
||||
\cref{tab:xray_eff}.
|
||||
%The absolute efficiencies for the $(2p-1s)$ lines of aluminium
|
||||
%(\SI{346.828}{\keV}) and silicon (\SI{400.177}{\keV})
|
||||
%are presented in \cref{tab:xray_eff}.
|
||||
In the process of efficiency calibration,
|
||||
corrections for true coincidence summing and self-absorption were not applied.
|
||||
The true coincidence summing probability is estimated to be very
|
||||
small, about \num{5.4d-6}, thanks to the far geometry of the calibration. The
|
||||
absorption in the source cover made of \SI{22}{\mg\per\cm^2}
|
||||
polyethylene is less than \num{4d-4} for a \SI{100}{\keV} photon.
|
||||
|
||||
A Monte Carlo (MC) study on the acceptance of the germanium detector with two
|
||||
purposes:
|
||||
Following corrections for the peak areas are considered:
|
||||
\begin{enumerate}
|
||||
\item compare between measured and MC efficiencies: a point source made of
|
||||
$^152$Eu is placed at the target position
|
||||
\item estimate the uncertainty due to finite-size geometry: the source is
|
||||
made of silicon with the same dimensions as those of the thick silicon
|
||||
detector, namely \SI[product-units=power]{1.5 x 50 x 50}{\mm}; then the
|
||||
primary vertex of $^152$Eu is generated inside the source.
|
||||
\item Correction for counting loss due to finite response time of the
|
||||
detector system, where two gamma rays arrive at the detector within a time
|
||||
interval short compared to that response time. This correction is
|
||||
significant in our germanium system because of the current pulse
|
||||
information extracting method does not count the second pulse.
|
||||
\item Correction of counting time loss in the reset periods of the transistor
|
||||
reset preamplifier. A preamplifier of this type would reset itself after
|
||||
accumulating a predetermined amount of charge. During a reset, the
|
||||
preamplifier is insensitive so this can be counted as dead time.
|
||||
\item True coincidence summing correction: two cascade gamma rays hit the
|
||||
detector at the same time would cause loss of count under the two
|
||||
respective peaks and gain under the sum energy peak.
|
||||
\item Correction for self-absorption of a gamma ray by the source itself.
|
||||
\end{enumerate}
|
||||
|
||||
The corrections for the first two mechanisms can be estimated by examining
|
||||
pulse length and intervals between two consecutive pulses in the germanium
|
||||
detector (\cref{fig:ge_cal_rate_pulselength}). The average pulse
|
||||
length is \SI{45.7}{\um}, the average count rate obtained from the decay rate
|
||||
of the interval spectrum is \SI{240}{\per\s}.
|
||||
|
||||
The correction factor for the finite response time of the detector system is
|
||||
calculated as:
|
||||
\begin{align}
|
||||
k_{\textrm{finite response time}} &= e^{2\times \textrm{(pulse length)}
|
||||
\times \textrm{(count rate)}}\\
|
||||
&= e^{2\times 47.5\times10^{-6} \times 241} \nonumber\\
|
||||
&= 1.02 \label{eqn:finite_time_response}
|
||||
\end{align}
|
||||
|
||||
The resets of the preamplifier show up as a peak around \SI{2}{\ms},
|
||||
consistent with specification of the manufacturer. Fitting the peak on top of
|
||||
an exponential background gives the actual reset pulse length of
|
||||
\SI{1947.34}{\us} and the number of resets during the calibration runs is
|
||||
2335.0. The total time loss for resetting is hence:
|
||||
$1947.34\times 10^{-6} \times 2335.0 = 4.55$ \si{\s}. That is a 0.14\% loss
|
||||
for a measuring time of \SI{3245.5}{\s}. This percentage loss is insignificant
|
||||
compared with the loss in \eqref{eqn:finite_time_response} and the statistical
|
||||
uncertainty of peak areas so correction for amplifier resets is not applied.
|
||||
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
\includegraphics[width=0.95\textwidth]{figs/ge_cal_rate_pulselength}
|
||||
\caption{Germanium detector pulse length (left) and intervals between pulses
|
||||
on that detector (right). The peak around \SI{2}{\ms} corresponds to the
|
||||
resets of the preamplifier. The peak at \SI{250}{\us} is due to triggering
|
||||
by the timing channel which is on the same digitiser.}
|
||||
\label{fig:ge_cal_rate_pulselength}
|
||||
\end{figure}
|
||||
|
||||
The true coincidence summing probability is estimated to be very small, about
|
||||
\num{5.4d-6}, thanks to the far geometry of the calibration. The absorption in
|
||||
the source cover made of \SI{22}{\mg\per\cm^2} polyethylene is less than
|
||||
\num{4d-4} for a \SI{100}{\keV} photon. Therefore these two corrections are
|
||||
omitted.
|
||||
|
||||
The absolute efficiencies of the reference gamma rays show agreement with those
|
||||
obtained from a Monte Carlo (MC) study where a point source made of $^{152}$Eu
|
||||
is placed at the target position (see \cref{fig:ge_eff_cal}). A comparison
|
||||
between efficiencies in case of the point-like source and a finite-size
|
||||
source is also done by MC simulation. As shown in \cref{fig:ge_eff_cal}, the
|
||||
differences are in line with the uncertainties of the measured efficiencies.
|
||||
%The dimensions of the latter are set to
|
||||
%resemble the distribution of muons inside the target: Gaussian spreading
|
||||
%\SI{11}{\mm} vertically, \SI{13}{\mm} horizontally, and \SI{127}{\um} in
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
\includegraphics[width=0.40\textwidth]{figs/ge_eff_cal}
|
||||
\includegraphics[width=0.40\textwidth]{figs/ge_eff_mc_finitesize_vs_pointlike_root}
|
||||
\caption{Absolute efficiency of the germanium detector, the fit was done with
|
||||
7 energy points from 244~keV, the shaded area is
|
||||
95\% confidence interval of the fit.}
|
||||
%because it is known that the linearity between
|
||||
%$ln(\textrm{E})$ and $ln(\textrm{eff})$ holds better.
|
||||
\label{fig:ge_eff_cal}
|
||||
\end{figure}
|
||||
The absolute efficiencies of the referenced points, and calculated efficiencies
|
||||
at X-rays of interest are listed in \cref{tab:xray_eff}.
|
||||
\begin{table}[htb]
|
||||
\begin{center}
|
||||
\pgfplotstabletypeset[
|
||||
@@ -601,18 +658,6 @@ purposes:
|
||||
\label{tab:xray_eff}
|
||||
\end{table}
|
||||
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
\includegraphics[width=0.40\textwidth]{figs/ge_eff_cal}
|
||||
\includegraphics[width=0.40\textwidth]{figs/ge_eff_mc_finitesize_vs_pointlike_root}
|
||||
\caption{Absolute efficiency of the germanium detector, the fit was done with
|
||||
7 energy points from 244~keV, the shaded area is
|
||||
95\% confidence interval of the fit.}
|
||||
%because it is known that the linearity between
|
||||
%$ln(\textrm{E})$ and $ln(\textrm{eff})$ holds better.
|
||||
\label{fig:ge_eff_cal}
|
||||
\end{figure}
|
||||
|
||||
% subsection germanium_detector (end)
|
||||
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
||||
%\subsection{Beam tuning and muon momentum scanning}
|
||||
@@ -684,12 +729,12 @@ different targets were carried out for silicon targets:
|
||||
|
||||
As the emitted protons deposit a significant amount of energy in the target
|
||||
material, thin targets and thus excellent momentum resolution of the low energy
|
||||
muon beam are critical. Aluminium targets of 50-\si{\micro\meter}\ and
|
||||
100~\si{\micro\meter}\ thick were used. Although a beam with low momentum spread of
|
||||
1\% is preferable, it was used for only a small portion of the run due to the
|
||||
low beam rate (see \cref{fig:Rates}). The beam momentum for each target
|
||||
was chosen to maximise the number of stopped muons. The collected data sets are
|
||||
shown in \cref{tb:stat}.
|
||||
muon beam are critical, aluminium targets of 50-\si{\micro\meter}\ and
|
||||
100-\si{\micro\meter}\ thick were used. Although a beam with low momentum
|
||||
spread of 1\% is preferable, it was used for only a small portion of the run
|
||||
due to the low beam rate (see \cref{fig:Rates}). The beam momentum for each
|
||||
target was chosen to maximise the number of stopped muons. The collected data
|
||||
sets are shown in \cref{tb:stat}.
|
||||
|
||||
\begin{table}[btp!]
|
||||
\begin{center}
|
||||
@@ -865,20 +910,8 @@ update the plots to reflect real-time status of the detector system.
|
||||
Some offline analysis modules has been developed during the beam time and could
|
||||
provide quick feedback in confirming and guiding the decisions at the time. For
|
||||
example, the X-ray spectrum analysis was done to confirm that we could observe
|
||||
the muon capture process (\cref{fig:muX}), and to help in choosing optimal
|
||||
momenta which maximised the number of stopped muons.
|
||||
\begin{figure}[btp]
|
||||
\centering
|
||||
\includegraphics[width=0.7\textwidth]{figs/muX.png}
|
||||
\caption{Germanium
|
||||
detector spectra in the range of 300 - 450 keV with different setups: no
|
||||
target, 62-\si{\micro\meter}-thick silicon target, and
|
||||
100-\si{\micro\meter}-thick aluminium target. The ($2p-1s$) lines from
|
||||
aluminium (346.828 keV) and silicon (400.177 keV) are clearly visible,
|
||||
the double peaks at 431 and 438 keV are from the lead shield, the peak at
|
||||
351~keV is a background gamma ray from $^{211}$Bi.}
|
||||
\label{fig:muX}
|
||||
\end{figure}
|
||||
the muon capture process and to help in choosing optimal momenta which
|
||||
maximised the number of stopped muons.
|
||||
|
||||
Although the offline analyser is still not fully developed yet, several modules
|
||||
are ready. They are described in detailed in the next chapter.
|
||||
|
||||
Reference in New Issue
Block a user