report for 1809 keV emission rate
This commit is contained in:
294
r15a_1809_rate/r15a_gamma.tex
Normal file
294
r15a_1809_rate/r15a_gamma.tex
Normal file
@@ -0,0 +1,294 @@
|
||||
\documentclass[11pt]{article}
|
||||
\usepackage{mhchem}
|
||||
\usepackage{hyperref}
|
||||
\usepackage{booktabs}
|
||||
\usepackage{multirow}
|
||||
\usepackage{textcomp}
|
||||
\usepackage{epsfig}
|
||||
\usepackage[noabbrev, capitalize]{cleveref}
|
||||
\usepackage[detect-weight=true, per=slash, detect-family=true, separate-uncertainty=true, alsoload=hep]{siunitx}
|
||||
|
||||
% \DeclareSIUnit\eVperc{\eV\per\clight}
|
||||
% \DeclareSIUnit\clight{\text{\ensuremath{c}}}
|
||||
|
||||
\begin{document}
|
||||
\title{Measurement of photons from nuclear muon capture on aluminum}
|
||||
\author{Nam H. Tran \\ Boston University}
|
||||
\date{\today}
|
||||
\maketitle
|
||||
|
||||
\begin{abstract}
|
||||
The goal is study the emission rate of the \SI{1808.7}{\kilo\eV} gamma rays
|
||||
(from the first excited state of \ce{^{26}Mg}) after nuclear muon capture on
|
||||
\ce{^{27}Al}. The measured emission rate is \SI{51(5)}{\percent} per muon capture.
|
||||
\end{abstract}
|
||||
|
||||
\section{Experimental set up}
|
||||
\label{sec:experimental_set_up}
|
||||
This measurement is part of the AlCap experiment done at PSI, Switzerland.
|
||||
The 2015 summer run focused on the detection of neutral particles: low energy
|
||||
X-ray, gamma ray and neutron emission after the muon is captured by the
|
||||
nucleus.
|
||||
|
||||
The X-rays and gamma rays of interest are:
|
||||
\begin{itemize}
|
||||
\item muonic $2p-1s$ transition in aluminum: \SI{346.8}{\kilo\eV}
|
||||
\item \SI{843.7}{\kilo\eV} gamma from the $\beta^-$ decay of \ce{^{27}Mg}
|
||||
(half-life: \SI{9.46}{\min})
|
||||
\item \SI{1808.7}{\kilo\eV} gamma from the first excited state of
|
||||
\ce{^{26}Mg}
|
||||
\end{itemize}
|
||||
|
||||
Low momentum muons (less than \SI[]{40}{\mega\eVperc}) were stopped in
|
||||
a target after passing a muon counter
|
||||
(\SI{60}{\mm}$\times$\SI{60}{\mm}$\times$\SI{0.5}{\mm} plastic scintillator).
|
||||
% Upstream from the muon counter, a
|
||||
% \SI{10}{\cm} $\times$ \SI{10}{\cm} $\times$ \SI{0.6}{\cm} scintillator with
|
||||
% a \SI{40}{\mm} diameter hole cut in the center acted as a beam defining
|
||||
% veto counter to the incoming muon beam.
|
||||
There were two 5"$\times$2" liquid scintillator BC501a detectors setup on the
|
||||
beam right to detect neutrons. For gamma spectrum analysis and normalization
|
||||
we used an HPGe detector installed on the beam left. In addition, a \ce{LaBr3}
|
||||
scintillator was also tested if it would be suitable to use in the STM. A 25
|
||||
LYSO crystal array was placed downstream of the target beam left to observe
|
||||
high energy photons emitted.
|
||||
|
||||
Two identical preamplifier outputs from the HPGe detector were fed into: (a)
|
||||
a timing filter amplifier for timing information, and (b) a spectroscopy
|
||||
amplifier for energy information. The timing pulses were read out by a 14-bit
|
||||
500-MS/s desktop digitizer(CAEN DT5730). In order to accommodate both low
|
||||
energy X-rays and relatively high energy gamma rays, we used two channels from
|
||||
the spectroscopy amplifier with different gain settings: (a) a lower gain
|
||||
channel for photons up to \SI{6.5}{\mega\eV}; and (b) a higher gain channel for
|
||||
photons up to \SI{2.5}{\mega\eV}. These channels were read out by a 14-bit
|
||||
100-MS/s VME digitizer (CAEN V1724).
|
||||
|
||||
The \ce{LaBr3} crystal is coupled with a photomultiplier, of which output
|
||||
pulses were large enough so no further amplification was needed. This channel
|
||||
is read out with the DT5730.
|
||||
|
||||
% Detectors' outputs were read out using waveform digitizers. We used a 14-bit
|
||||
% 100-MS/s VME digitizer (CAEN V1724) to record energy signals from
|
||||
% HPGe and \ce{LaBr3} detectors. There were two energy outputs from the HPGe
|
||||
% detector with different gain settings: (a) low gain channel for photons up to
|
||||
% \SI{6.5}{\mega\eV}; and (b) high gain channel for photons up to
|
||||
% \SI{2.5}{\mega\eV}. The timing signals from these detectors, and signals from
|
||||
% plastic and liquid scintillators were fed into a faster digitizer, namely
|
||||
% a 14-bit 500-MS/s desktop digitizer (CAEN DT5730).
|
||||
% These fast timing channels
|
||||
% were also read out using a multihit TDC (CAEN V1290A) as a back up solution.
|
||||
% All digitizers and TDC were synchronized by an external master clock.
|
||||
|
||||
Experimental layout is shown in \cref{fig:R2015a_setup}.
|
||||
\begin{center}
|
||||
\begin{figure}[!tbp]
|
||||
\centering
|
||||
\includegraphics[width=0.70\textwidth]{figs/R2015a_setup_2.jpg}
|
||||
\caption{Layout of the AlCap 2015 summer run. Muons entered from the top of
|
||||
the image. The LYSO detector is not visible in this image, which is
|
||||
located further out in the bottom of the image.}
|
||||
\label{fig:R2015a_setup}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
|
||||
There were several runs with different targets made of aluminum, titanium,
|
||||
lead, water. All targets were sufficiently thick to stop the muon beam with
|
||||
momenta up to \SI{40}{\mega\eVperc}.
|
||||
% Table~\ref{tab:alcap2015a:datasets} summarizes the data
|
||||
% sets for the main production run and number of muons
|
||||
% entering the experiment as counted by the beam scintillator counter TSc. Data
|
||||
% were also collected on stainless steel, tungsten and mylar targets with
|
||||
% a substantially reduced amount of data collection time.
|
||||
|
||||
\section{Analysis}
|
||||
\label{sec:analysis}
|
||||
In this study, the dataset on a \SI{2}{\mm} thick aluminum target is used. It
|
||||
was collected in 31 hours of beam time and contains about \num{1.91E9} stopped
|
||||
muons. Momentum of the muon beam was \SI{36}{\mega\eVperc}.
|
||||
|
||||
\subsection{Digital pulse processing}
|
||||
\label{sub:digital_pulse_processing}
|
||||
Since we recorded all detector outputs using digitizers, offline digital pulse
|
||||
processing is needed to extract energy and timing information. Typical output
|
||||
pulses from HPGe and \ce{LaBr3} detectors are shown in
|
||||
\cref{fig:typical_pulses}.
|
||||
|
||||
\begin{center}
|
||||
\begin{figure}[htbp]
|
||||
\centering
|
||||
\includegraphics[width=1.0\textwidth]{figs/typical_pulses}
|
||||
\caption{Typical output pulses of HPGe and \ce{LaBr3} detectors: energy
|
||||
output HPGe high gain (top left), energy output HPGe low gain (top
|
||||
right), timing output HPGe (bottom left), and \ce{LaBr3} (bottom right).}
|
||||
\label{fig:typical_pulses}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
|
||||
The timing pulses from the HPGe detector were not used in this analysis because
|
||||
they are too long and noisy (see bottom left \cref{fig:typical_pulses}).
|
||||
Energy of the HPGe detector is taken as amplitude of spectroscopy amplifier
|
||||
outputs, its timing is determined by the clock tick where the trace passing
|
||||
\SI{30}{\percent} of the amplitude.
|
||||
|
||||
\ce{LaBr3} pulses were passed through a moving average window filter (60
|
||||
samples wide), then integrated to obtain energy resolution.
|
||||
|
||||
\subsection{Calibrations}
|
||||
\label{sub:calibrations}
|
||||
The HPGe and \ce{LaBr3} detectors acceptance and energy scales were calibrated
|
||||
using \ce{^{152}Eu}, \ce{^{60}Co}, \ce{^{88}Y} sources placed at the target
|
||||
position. There was a separate run for background radiation.
|
||||
|
||||
\cref{fig:uncalibrated_labr3_spectra} shows \ce{LaBr3}
|
||||
spectra with calibration sources \ce{^{88}Y}, \ce{^{60}Co}, and background
|
||||
radiation. It can be seen that the self activation from \ce{Ac} dominates the
|
||||
spectra. The \SI{1173}{\kilo\eV} peak barely shows up in \ce{^{60}Co}
|
||||
spectrum, while the \SI{1332}{\keV} peak is buried under the
|
||||
\SI{1436}{\kilo\eV} peak from \ce{^{138}La}. The \SI{1836}{\kilo\eV}
|
||||
peak of \ce{^{88}Y} and the annihilation peak \SI{511}{\kilo\eV} can be
|
||||
distinguished, but the \SI{898}{\kilo\eV} has been distorted by the electrons
|
||||
and \SI{789}{\kilo\eV} gammas from the beta decay of \ce{^{138}La}. The energy
|
||||
resolution (FWHM) at the \SI{1836}{\kilo\eV} peak was \SI{5.9}{\percent}.
|
||||
|
||||
\begin{center}
|
||||
\begin{figure}[htbp]
|
||||
\centering
|
||||
\includegraphics[width=1.0\textwidth]{figs/labr3_spectra_w_gatedintegration}
|
||||
\caption{Calibration of the \ce{LaBr3} detector, top horizontal axis shows
|
||||
energy and bottom horizontal axis shows integration of the output pulses.
|
||||
The spectra were scaled to make the peak recognition easier.}
|
||||
\label{fig:uncalibrated_labr3_spectra}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
|
||||
The HPGe spectra are much cleaner as shown in Figure~\ref{fig:hpge_ecal}.
|
||||
\begin{center}
|
||||
\begin{figure}[htbp]
|
||||
\centering
|
||||
\includegraphics[width=1.0\textwidth]{figs/hpge_ecal}
|
||||
\caption{Energy calibration spectra for the HPGe detector.}
|
||||
\label{fig:hpge_ecal}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
% Energy resolutions were good for all calibration peaks.
|
||||
The detector acceptance
|
||||
were fitted as a function of photon energy above \SI{200}{\kilo\eV}:
|
||||
\begin{equation}
|
||||
A = c_1 \times E ^ {c_2},
|
||||
\end{equation}
|
||||
where $c_1 = 0.1631$, $c_2 = -0.9257$. Interpolation gives detector acceptance
|
||||
at the peaks of interest as shown in \cref{tab:hpge_acceptance}.
|
||||
|
||||
\begin{center}
|
||||
\begin{figure}[htbp]
|
||||
\centering
|
||||
\includegraphics[width=1.0\textwidth]{figs/hpge_higain_acceptance}
|
||||
\caption{Acceptance of the HPGe (high gain channel) as a function of photon
|
||||
energy.}
|
||||
\label{fig:hpge_higain_acceptance}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
|
||||
\begin{table}[htbp]
|
||||
\centering
|
||||
\caption{HPGe acceptance for photons of interest}
|
||||
\label{tab:hpge_acceptance}
|
||||
\begin{tabular}{@{}cccc@{}}
|
||||
\toprule
|
||||
\multicolumn{2}{c}{\textbf{\begin{tabular}[c]{@{}c@{}}Photon energy\\ {[}keV{]}\end{tabular}}} & \textbf{Acceptance} & \textbf{Error} \\
|
||||
\midrule
|
||||
$2p-1s$ & 346.8 & \num{7.26e-4} &\num{4.73e-5} \\
|
||||
% 3p-1s & 399.3 & \num{6.38e-4} &\num{3.71e-5} \\
|
||||
% 4p-1s & 400.2 & \num{6.36e-4} &\num{3.70e-5} \\
|
||||
% 5p-1s & 476.8 & \num{5.41e-4} &\num{2.72e-5} \\
|
||||
\ce{^{27}Mg} & 843.7 & \num{3.19e-4} &\num{1.20e-5} \\
|
||||
% & 1014.4 & \num{2.69e-4} &\num{1.07e-5} \\
|
||||
\ce{^{26}Mg}* & 1088.7 & \num{1.57e-4} &\num{9.80e-6} \\
|
||||
\bottomrule
|
||||
\end{tabular}
|
||||
\end{table}
|
||||
|
||||
\section{Results and discussion}
|
||||
\label{sec:results_and_discussion}
|
||||
|
||||
\subsection{\ce{LaBr3} spectra}
|
||||
\label{sub:labr3_spectra}
|
||||
The \ce{LaBr3} energy spectra for the Al dataset are presented in
|
||||
\cref{fig:labr3_all_al_runs}. The muonic $2p-1s$ peak shows up clearly in
|
||||
the prompt spectrum as expected. The \SI{1809}{\kilo\eV} peak can be
|
||||
recognized, it has better
|
||||
signal-to-background ratio in the prompt spectrum than in the delay spectrum
|
||||
(0.88 to 0.33). The background under the \SI{1809}{\kilo\eV} is dominated by
|
||||
the $\alpha$ decay of progenies from \ce{^{227}Ac}. I think that this
|
||||
\ce{LaBr3} in the current set up is not suitable to use as a STM detector.
|
||||
\begin{center}
|
||||
\begin{figure}[htbp]
|
||||
\centering
|
||||
\includegraphics[width=1.0\textwidth]{figs/labr3_all_al_runs}
|
||||
\caption{\ce{LaBr3} spectra: prompt (less than \SI{100}{\ns} from muon
|
||||
hit), delay ($>$ \SI{100}{\ns} from muon hit), and all hits.}
|
||||
\label{fig:labr3_all_al_runs}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
|
||||
\subsection{HPGe spectrum}
|
||||
\label{sec:hpge_spectrum}
|
||||
The HPGe photon spectrum for the aluminum dataset is shown in
|
||||
\cref{fig:GeCHH_all_al_runs}. Both the \SI{347}{\kilo\eV}
|
||||
and \SI{1809}{\keV} peaks are clearly visible with the X-ray peak dominates in
|
||||
the prompt spectrum. The apperance of the \SI{347}{\keV} (and other X-ray
|
||||
peaks) in the delay spectrum can be explained by a second muon stopped in the
|
||||
aluminum target shortly after the trigger muon.
|
||||
\begin{center}
|
||||
\begin{figure}[htbp]
|
||||
\centering
|
||||
\includegraphics[width=1.0\textwidth]{figs/GeCHH_all_al_runs}
|
||||
\caption{HPGe high gain spectra: prompt (less than \SI{500}{\ns} from muon
|
||||
hit), delay ($>$ \SI{500}{\ns} from muon hit), and all hits.}
|
||||
\label{fig:GeCHH_all_al_runs}
|
||||
\end{figure}
|
||||
\end{center}
|
||||
|
||||
\subsection{Number of stopped muons}
|
||||
\label{sub:number_of_stopped_muons}
|
||||
|
||||
The number of stopped muons is calculated by two methods:
|
||||
\begin{itemize}
|
||||
\item infering from the number of $2p-1s$ X-rays,
|
||||
\item counting the muon hits on the muon counter.
|
||||
\end{itemize}
|
||||
The latter gives the number of stopped muons as:
|
||||
\begin{equation}
|
||||
N_{\mu} = 3.03 \times 10^8 \pm 1.7 \times 10^4.
|
||||
\label{eqn:n_mu_TSc}
|
||||
\end{equation}
|
||||
|
||||
The number of $2p-1s$ X-rays is calculated by fitting a Gaussian peak with
|
||||
a linear background to the region \SIrange{340}{350}{\keV} around the peak in
|
||||
the prompt HPGe spectrum:
|
||||
\begin{equation}
|
||||
N_{346.8} = (191.27 \pm 0.42) \times 10^3.
|
||||
\end{equation}
|
||||
Using the acceptance of the $2p-1s$ photons in \cref{tab:hpge_acceptance},
|
||||
number of stopped muons is:
|
||||
\begin{equation}
|
||||
N_{\mu} = \frac{N_{346.8}}{A_{346.8}} = (3.30 \pm 0.22) \times 10^8,
|
||||
\end{equation}
|
||||
which is consistent with that in \cref{eqn:n_mu_TSc}.
|
||||
|
||||
\subsection{Emission rate of \SI{1809}{\keV} photons}
|
||||
\label{sub:emission_rate_of_1809_kev_photons}
|
||||
Number of the \SI{1809}{\keV} photons is calculated using the same method for
|
||||
the \SI{347}{\keV} photons:
|
||||
\begin{equation}
|
||||
N_{1808.7} = 16032.54 \pm 166.19.
|
||||
\end{equation}
|
||||
Therefore the emission rate per nuclear capture is:
|
||||
\begin{equation}
|
||||
R_{1808.7} = \frac{N_{1808.7}}{A_{1808.7} \times N_{\mu} \times 0.609} = 0.51 \pm 0.05,
|
||||
\end{equation}
|
||||
, where the factor 0.609 comes from the fact that only \SI{60.9}{\percent} of
|
||||
stopped muons are captured. This result is consistent with the rate reported
|
||||
by Measday et al.
|
||||
|
||||
\end{document}
|
||||
Reference in New Issue
Block a user