prog saved
This commit is contained in:
@@ -2,9 +2,8 @@
|
||||
\label{cha:the_alcap_run_2013}
|
||||
\thispagestyle{empty}
|
||||
The first run of the AlCap experiment was performed at the $\pi$E1 beam line
|
||||
area, PSI from November 26 to December 23,
|
||||
2013. The goal of the run was to measure protons rate and spectrum following
|
||||
muon capture on aluminium.
|
||||
area, PSI from November 26 to December 23, 2013. The goal of the run was to
|
||||
measure protons rate and their spectrum following muon capture on aluminium.
|
||||
|
||||
\section{Experimental set up}
|
||||
\label{sec:experimental_set_up}
|
||||
@@ -56,7 +55,7 @@ One of the main requirements of the AlCap experiment was a low energy muon beam
|
||||
with narrow momentum bite in order to achieve a high fraction of stopping muons
|
||||
in the very thin targets. In this Run 2013, muons from
|
||||
\SIrange{28}{45}{\mega\electronvolt\per\cc} and momentum spread of 1\% and
|
||||
3\%were used.
|
||||
3\%, respectively, were used.
|
||||
|
||||
For part of the experiment the target was replaced with one of the silicon
|
||||
detector packages allowed an accurate momentum and range calibration
|
||||
@@ -84,8 +83,9 @@ as a function of momentum for two different momentum bites.
|
||||
|
||||
The targets and charged particle detectors are installed inside the vacuum
|
||||
chamber as shown in \cref{fig:alcap_setup_detailed}. The muon beam enters
|
||||
from the right of the image and hits the target, which is placed at the
|
||||
centre of the vacuum chamber and orientated at 45 degrees to the beam axis.
|
||||
from the right of \cref{fig:alcap_setup_detailed} and hits the target, which is
|
||||
placed at the centre of the vacuum chamber and orientated at 45 degrees to the
|
||||
beam axis.
|
||||
The side walls and bottom flange of the vessel provide several
|
||||
vacuum-feedthroughs for the high voltage and signal cables for the silicon and
|
||||
scintillator detectors inside the chamber.
|
||||
@@ -207,7 +207,7 @@ proportional chamber (\Pmu{}PC) that uses 24 X wires and 24 Y wires at
|
||||
2~\si{\milli\meter}~intervals.
|
||||
|
||||
The upstream detectors provide signal of an incoming muon as coincident hits on
|
||||
the muon trigger and the wire chamber in anti-coincident with the muon
|
||||
the muon trigger and the wire chamber in anti-coincidence with the muon
|
||||
anti-coincidence counter.
|
||||
This set of detectors along with their read-out system
|
||||
belong to the MuSun experiment, which operated at the same beam line just
|
||||
@@ -326,7 +326,7 @@ lead to the use of several sampling frequencies from 17~\si{\mega\hertz}\ to
|
||||
250~\si{\mega\hertz}, and three types of digitisers were employed:
|
||||
\begin{itemize}
|
||||
\item custom-built 12-bit 170-MHz FADCs which was designed for the
|
||||
MuCap experiment. Each FADC board has dimensions the same as those of
|
||||
MuCap experiment. Each FADC board has the same dimensions as those of
|
||||
a single-width 6U VME module, but is hosted in a custom built crate due to
|
||||
its different power supply mechanical structure. The FADC communicates with
|
||||
a host computer through a 100-Mb/s Ethernet interface using a simple
|
||||
@@ -367,7 +367,7 @@ model V767 which was kindly provided by the MuSun experiment.
|
||||
The Data Acquisition System (DAQ) of the AlCap experiment, so-called AlCapDAQ,
|
||||
provided the readout of front-end electronics, event assembling, data logging,
|
||||
hardware monitoring and control, and the run database of the experiment
|
||||
(\cref{fig:alcapdaq_pcs}). It was based on MIDAS framework~\footnote{
|
||||
(\cref{fig:alcapdaq_pcs}). It was based on the MIDAS framework~\footnote{
|
||||
MIDAS is a general purpose DAQ software system developed at PSI and TRIUMF:\\
|
||||
\url{http://midas.triumf.ca}} and consisted of two circuits, {\em i})
|
||||
a detector circuit for synchronous data readout from the front-end electronics
|
||||
@@ -411,7 +411,7 @@ rollover on any digitiser (a FADC runs at its maximum speed of
|
||||
To ease the task of handling data, the data collecting period was divided into
|
||||
short runs, each run stopped when the logger had recorded 2 GB of data.
|
||||
The data size effectively made each run last for about 5 minutes. The DAQ
|
||||
automatically starts a new run with the same parameters after about 6 seconds.
|
||||
automatically started a new run with the same parameters after about 6 seconds.
|
||||
The short period of each run also allows the detection, and helps to reduce the
|
||||
influence of effects such as electronics drifting, temperature fluctuation.
|
||||
|
||||
@@ -435,11 +435,12 @@ run, by:
|
||||
would lose about \SI{66}{\kilo\eV} in the \SI{0.5}{\um}-thick dead layer,
|
||||
and the peak would appear at \SI{5418}{\kilo\eV} (\cref{fig:toyMC_alpha});
|
||||
|
||||
\item and a tail pulse generator, A tail pulse with amplitude of
|
||||
\item a tail pulse generator, A tail pulse with amplitude of
|
||||
\SI{66}{\milli\volt}~was used to simulate the response of the silicon
|
||||
detectors' preamplifiers to a particle with \SI{1}{\MeV} energy deposition;
|
||||
detectors' preamplifiers to a particle with \SI{1}{\MeV} energy
|
||||
deposition; and
|
||||
|
||||
\item During data taking period, electrons in the beam were were also used
|
||||
\item during data taking period, electrons in the beam were were also used
|
||||
for energy calibration of thick silicon detectors where energy deposition
|
||||
is large enough. The muons at different momenta provided another mean of
|
||||
calibration in the beam tuning period.
|
||||
@@ -453,6 +454,10 @@ run, by:
|
||||
\label{fig:toyMC_alpha}
|
||||
\end{figure}
|
||||
|
||||
The conversion from ADC value to energy is done with a first-order polynomial:
|
||||
\begin{equation}
|
||||
\textrm{E [keV]} = \textrm{Slope} \times \textrm{ADC} + \textrm{Offset}.
|
||||
\end{equation}
|
||||
The calibration coefficients for the silicon channels are listed in
|
||||
\cref{tab:cal_coeff}.
|
||||
|
||||
@@ -472,6 +477,11 @@ The calibration coefficients for the silicon channels are listed in
|
||||
every head row/.style={
|
||||
before row={\toprule},
|
||||
after row={\midrule},
|
||||
%%TODO unit of coeffcients
|
||||
%after row={ \arraybackslash
|
||||
%{ }& { keV/channel } & { keV }\\
|
||||
%\midrule},
|
||||
%{}& {(keV/channel)} & {(keV)}\\ \midrule},
|
||||
columns/Detector/.style={column type=c},
|
||||
columns/Slope/.style={column type=c},
|
||||
columns/Offset/.style={column type=c}
|
||||
@@ -534,13 +544,14 @@ Following corrections for the peak areas are considered:
|
||||
detector system, where two gamma rays arrive at the detector within a time
|
||||
interval short compared to that response time. This correction is
|
||||
significant in our germanium system because of the current pulse
|
||||
information extracting method does not count the second pulse.
|
||||
information extracting method does not count the second pulse (see
|
||||
\cref{sub:offline_analyser}).
|
||||
\item Correction of counting time loss in the reset periods of the transistor
|
||||
reset preamplifier. A preamplifier of this type would reset itself after
|
||||
accumulating a predetermined amount of charge. During a reset, the
|
||||
preamplifier is insensitive so this can be counted as dead time.
|
||||
preamplifier is insensitive so this can be counted as a dead time.
|
||||
\item True coincidence summing correction: two cascade gamma rays hit the
|
||||
detector at the same time would cause loss of count under the two
|
||||
detector at the same time would cause loss of counts under the two
|
||||
respective peaks and gain under the sum energy peak.
|
||||
\item Correction for self-absorption of a gamma ray by the source itself.
|
||||
\end{enumerate}
|
||||
@@ -568,7 +579,7 @@ an exponential background gives the actual reset pulse length of
|
||||
$1947.34\times 10^{-6} \times 2335.0 = 4.55$ \si{\s}. That is a 0.14\% loss
|
||||
for a measuring time of \SI{3245.5}{\s}. This percentage loss is insignificant
|
||||
compared with the loss in \eqref{eqn:finite_time_response} and the statistical
|
||||
uncertainty of peak areas so correction for amplifier resets is not applied.
|
||||
uncertainty of peak areas.
|
||||
|
||||
\begin{figure}[htb]
|
||||
\centering
|
||||
@@ -590,8 +601,12 @@ The absolute efficiencies of the reference gamma rays show agreement with those
|
||||
obtained from a Monte Carlo (MC) study where a point source made of $^{152}$Eu
|
||||
is placed at the target position (see \cref{fig:ge_eff_cal}). A comparison
|
||||
between efficiencies in case of the point-like source and a finite-size
|
||||
source is also done by MC simulation. As shown in \cref{fig:ge_eff_cal}, the
|
||||
differences are in line with the uncertainties of the measured efficiencies.
|
||||
source is also done by MC simulation. The differences between the two sources
|
||||
are generally smaller than 3\%, which are comparable with the uncertainties of
|
||||
the efficiency calibration. That means the point-like efficiencies can be used
|
||||
for a finite-sized source without correction.
|
||||
%As shown in \cref{fig:ge_eff_cal}, the
|
||||
%differences are in line with the uncertainties of the measured efficiencies.
|
||||
%The dimensions of the latter are set to
|
||||
%resemble the distribution of muons inside the target: Gaussian spreading
|
||||
%\SI{11}{\mm} vertically, \SI{13}{\mm} horizontally, and \SI{127}{\um} in
|
||||
@@ -599,9 +614,12 @@ differences are in line with the uncertainties of the measured efficiencies.
|
||||
\centering
|
||||
\includegraphics[width=0.40\textwidth]{figs/ge_eff_cal}
|
||||
\includegraphics[width=0.40\textwidth]{figs/ge_eff_mc_finitesize_vs_pointlike_root}
|
||||
\caption{Absolute efficiency of the germanium detector, the fit was done with
|
||||
7 energy points from 244~keV, the shaded area is
|
||||
95\% confidence interval of the fit.}
|
||||
\caption{Absolute efficiency of the germanium detector (right) and
|
||||
MC comparison of efficiencies in case of point-like and finite-sized
|
||||
sources (left). The efficiencies curve is fitted on
|
||||
7 measured energy points from \SIrange{244}{1408}{\keV}, the shaded area is
|
||||
95\% confidence interval of the fit. The ratios on the left plot are
|
||||
normalised to the efficiencies of the point-like case at each energy point.}
|
||||
%because it is known that the linearity between
|
||||
%$ln(\textrm{E})$ and $ln(\textrm{eff})$ holds better.
|
||||
\label{fig:ge_eff_cal}
|
||||
@@ -770,8 +788,8 @@ sets are shown in \cref{tb:stat}.
|
||||
\label{sub:concept}
|
||||
Since the AlCapDAQ is a trigger-less system, it stored all waveforms of the
|
||||
hits occured in 100-ms-long blocks without considering their physics
|
||||
significance The analysis code therefore must be able to extract parameters of
|
||||
the waveforms, then organises the pulses into physics events correlated to
|
||||
significance. The analysis code therefore must be able to extract parameters of
|
||||
the waveforms, then organises the pulses into the physics events correlated to
|
||||
stopped muons (\cref{fig:muon_event}). In addition, the analyser is
|
||||
intended to be usable as a real-time component of a MIDAS DAQ, where simple
|
||||
analysis could be done online for monitoring and diagnostic during the run.
|
||||
@@ -786,9 +804,9 @@ analysis could be done online for monitoring and diagnostic during the run.
|
||||
|
||||
The analysis framework of the AlCap consists of two separate programs.
|
||||
A MIDAS-based analyser framework, \alcapana{}, processes the raw data and
|
||||
passes its ROOT data output to a second
|
||||
passes its ROOT data output to the second
|
||||
stage, \rootana{}, where most of the physics analysis is performed.
|
||||
Both programs were designed to be modularised, which allowed us to develop
|
||||
Both of the programs were designed to be modularised, which allowed us to develop
|
||||
lightweight analysis modules that were used online to generate plots quickly,
|
||||
while more sophisticated modules can be applied in offline analysis.
|
||||
|
||||
@@ -908,14 +926,14 @@ update the plots to reflect real-time status of the detector system.
|
||||
%\hl{Screen shots}
|
||||
\subsection{Offline analyser}
|
||||
\label{sub:offline_analyser}
|
||||
Some offline analysis modules has been developed during the beam time and could
|
||||
Some offline analysis modules have been developed during the beam time and could
|
||||
provide quick feedback in confirming and guiding the decisions at the time. For
|
||||
example, the X-ray spectrum analysis was done to confirm that we could observe
|
||||
the muon capture process and to help in choosing optimal momenta which
|
||||
maximised the number of stopped muons.
|
||||
|
||||
Although the offline analyser is still not fully available yet, several modules
|
||||
are ready(\cref{tab:offline_modules}). An initial analysis is possible using
|
||||
are ready (\cref{tab:offline_modules}). An initial analysis is possible using
|
||||
the existing modules thanks to the modularity of the analysis framework.
|
||||
|
||||
\begin{table}[htb]
|
||||
@@ -1000,7 +1018,7 @@ the beam rate was generally less than \SI{8}{\kilo\hertz}.
|
||||
%\end{figure}
|
||||
|
||||
To make sure that we will analyse good data, a low level data quality checking
|
||||
was done on the whole data sets. The idea is plotting the variations of basic
|
||||
was done on the whole data sets. The idea is to plot the variations of basic
|
||||
parameters, such as noise level, length of raw waveforms, pulse rate, time
|
||||
correlation to hits on the muon counter on each channel during the data
|
||||
collecting period. Runs with significant difference from the averaging
|
||||
@@ -1026,13 +1044,13 @@ shown in \cref{fig:lldq}.
|
||||
|
||||
A full Monte Carlo (MC) simulation of the experimental set up has been developed
|
||||
based on Geant4~\cite{Agostinelli.etal.2003}. The geometrical implementation
|
||||
was as detailed as possible and could be modified via configuration script at
|
||||
run time. Descriptions of the muon beam came from the beam line optic
|
||||
was detailed as much as possible and could be modified via configuration
|
||||
scripts at run time. Descriptions of the muon beam came from the beam line optic
|
||||
calculation provided by the accelerator experts at PSI.
|
||||
|
||||
The MC model greatly assisted the design of the experiment, such as alignment
|
||||
of the detectors with respect to the target, and shielding of scattered muons.
|
||||
It also helps make sense of observed results during the run and data
|
||||
It also helped make a sense of the observed results during the run and data
|
||||
analysing.
|
||||
|
||||
% chapter the_alcap_run_2013 (end)
|
||||
|
||||
Reference in New Issue
Block a user